Panametrics OXY.IQ-031-10 Transmitter
Test Cert included
OXY.IQ: oxy.IQ Oxygen Transmitter; 4 to 20 mA output
0: No sensor
3: Intrinsically safe , US/CAN Class 1, Div 1, Groups ABCD, T4; ATEX / IECEx Ex ia IIC Ga T4
1: 2mtrs of cable
1: Zener Barrier
0: No Special Requirements
Applications An oxygen transmitter for use in: • Glove box purge and leak detection • Natural gas • Semiconductor wafer machines • Coating process machines • Membrane air separators • Inert welding gases • Pure gaseous hydrocarbon streams • Process monitoring of gaseous monomers • Heat treating and bright annealing Features • Two-wire, loop-powered 4 to 20 mA transmitter • Display with keypad • Intrinsically safe option (pending) • Proven galvanic fuel cell O2 sensor technology • User selectable ranges for ppm and percent oxygen • User-friendly and intuitive user interface with diagnostics • Microprocessor-based, all-digital technology for reliable operation • Low maintenance, economical and compact • Sensor failure output error • Sensor lifetime indication • NAMUR error indication
Oxygen Sensor Interference Gases |
|
|
|
H2S |
<5 ppm |
<10 ppm |
0.0005% |
0.01% |
0.001% |
0.1% |
SO3 |
<10 ppm |
<10 ppm |
0.01% |
0.1% |
0.01% |
0.1% |
SO2 |
<10 ppm |
(3) |
0.01% |
0.1% |
(3) |
(3) |
HCI |
<1000 ppm |
(3) |
0.1% |
1.0% |
(3) |
(3) |
HCN |
<1000 ppm |
(3) |
0.1% |
1.0% |
(3) |
(3) |
CO2 |
<1000 ppm |
(3) |
0.1% |
20% |
(3) |
(3) |
NO2 |
(2) |
(2) |
(2) |
(2) |
(2) |
(2) |
CL2 |
(2) |
(2) |
(2) |
(2) |
(2) |
(2) |
Cont. = Continuous, Int. = Intermittent |
|
|
|
(1) Recommended maximum exposure 30 minutes followed by flushing with |
ambient air for equal period |
|
|
|
|
(2) Minimal effect on sensor performance, but produces signal interference of |
1:2 ratio, ppm levels only. For example, 100 ppm NO2 looks like 200 ppm O2 |
(3) Minimal effect on sensor performance |
|
|
|
Power Requirements 24 to 28 VDC at 50 mA Cable • PN 704-1318: blue jacket, twisted pair with connector; 26 AWG; 2 or 10 m length Output Total load must equal 250 W ±5% when using zener barrier Non-Incendive (Div 2) and General Purpose Installation No zener barrier is used. Power Requirements 9 to 28 VDC loop powered, 0.7 W max Cable • PN 704-1317: black jacket, twisted pair with connector; 26 AWG; 2 or 10 m length All Installations Process Wetted Materials • SS process unit: 316 stainless steel, Viton® O-ring, gold plated sensor electrical contacts and glass User-Selectable Measurement Ranges • PPM sensors: �?0 to 10 ppmv O2 (OX-1 or OX-2 only) �?0 to 20 ppmv O2 (OX-1 or OX-2 only) �?0 to 50 ppmv O2 (OX-1 or OX-2 only) �?0 to 100 ppmv O2 �?0 to 200 ppmv O2 �?0 to 500 ppmv O2 �?0 to 1000 ppmv O2 �?0 to 2000 ppmv O2 �?0 to 5000 ppmv O2 �?0 to 10,000 ppmv O2 • Percent sensors: �?0% to 1% O2 �?0% to 2% O2 �?0% to 5% O2 �?0% to 10% O2 �?0% to 25% O2 �?0% to 50% O2 Accuracy • ±1% of range at calibration point • ±2% of range at the calibration point for the 0 to 10 ppmv range (OX-1 or OX-2 only) Repeatability • ±1% of range • ±2% of Range for the 0 to 10 ppmv range (OX-1, 2 only) Resolution ±0.1% of range Linearity ±2% of range (OX-1, 2, 3, 5) ±5% of range (OX-4) O2 Sensor Operating Temperature 32°F to 113°F (0°C to 45°C) Sample Pressure Vented to atmosphere during operation and calibration Atmospheric Pressure Effect ±0.13% of reading per mmHg (directly proportional to absolute pressure). During calibration, pressure and flow must be kept constant. Process Connection • 1/8 in NPT inlet and outlet Dimensions 4.10 in x 2.75 in x 2.05 in Weight 1.35 lb (612 grams)
Sample Flow Rate 1.0 SCFH (500 cc/min) recommended for process units Electrical Classification/Certification (Pending) Intrinsically Safe package with zener barrier USA/Canada IS for Class I, II, III, Div 1, Groups A, B, C, D, E, F, G, T6 AEx ia IIC T6; Ex ia IIC T6; Tamb -20 to +60°C EU ATEX and International IECEx Ex Ia IIC Ge T6 Ex ta IIC De T85C Tamb -20 to 60°C Standard package; Non-incendive without use of zener barrier or galvanic isolator USA/Canada Class I, Division 2, Groups A, B, C, D, T6 ATEX/ IECEx Ex na IIC T6 Weatherproof/Corrosion Resistant Type 4X IP66 European Compliance Complies with EMC Directive 2004/108/EC |