DM5E Series Wall Thickness Gauge - GE Measurement
DM5E Ultrasonic Thickness Gage
The DM5E Series Ultrasonic Wall Thickness Gage enables you to accomplish a large number of demanding tasks, especially with remaining wall thickness measurements on components subjected to corrosion and erosion. These gauges are smart, simple-to-use ultrasonic thickness gages that offer big performance in a small lightweight package. Their compact size and ergonomic shape provide comfortable one-hand use.
Features and Benefits
BENEFITS
Three levels - DM5E Basic, DM5E and DM5E DL
Thickness measurement under coating
User-friendly operator interface
New range of high performance probes
Flexible data processing
A range of measurement displays
FEATURES
Range of measurement displays:
Normal: the thickness value appears as large digits in the centre of the display
MIN Scan: a minimum thickness scan that allows the user to run the probe over the wall surface. After the evaluation period, the minimum material thickness measured is displayed
MAX Scan: a maximum thickness scan is exactly the same as a MIN Scan apart from the fact that the maximum thickness measured is displayed
DIFF/RR%: compares the measured thickness with a user-specified nominal thickness - dimensional difference between the two values is displayed, as is the percentage difference
B-Scan: displays a graphic representation of a B-Scan showing minimum thickness values - graph is derived by measuring and recording at 1 point per second
- See more at: http://0s.o53xo.m5sw2zlbon2xezlnmvxhiltdn5wq.erenta.ru/inspection-ndt/ultrasound/dm5e-series-wall-thickness-gauge#sthash.1fQM2REO.dpuf
Operating Principle |
Ultrasonic, Pulse Echo Measurement Method |
Measuring Range |
Depends on Probe and Material, 0.60 mm to 508 mm (0.025" to 20.00") in IP to 1st BW Measurement Mode, 2.00 mm to 127.0 mm (0.079" to 5.00")in Dual-Multi Measurement Mode, the Coating Thickness Range Shall be 0.3 mm to 2.50 mm (0" to 0.098"). |
Measuring Resolution |
0.01 mm Default - Selectable 0.01, 0.1 mm (0.001" Default - Selectable 0.001,0.01") |
Material Velocity Range |
0.508 to 18.699 mm/msec (0.0200 to 0.7362"/msec) |
Material Velocity Resolution |
1 m/s (0.0001"/msec) |
Units |
Inch or Millimeter |
Calibration |
One-Point Cal, On-Block and Off-Block, Two-Point Cal |
Pulser |
Excitation Pulse |
Spike Pulser |
Voltage |
120 V into 50 ohm Load, Using 20 MHz Oscilloscope |
Receiver |
Bandwidth |
500KHz to 12 MHz @ -3 dB |
Gain |
Automatic Gain Control |
Display Type |
High-Resolution Graphical LCD, 64 x 128 Pixels, 53.0 mm x 27.0 mm with Backlight and Adjustable Contrast |
Update Rate |
4 Hz or 8 Hz, User Selectable, 24 Hz Scan Mode Capture Rate |
Thickness Value Display |
NORMAL View Mode |
5 Digit, 10.6 mm (0.4") High |
B-SCAN View Mode |
5 Digit, 2.55 mm (0.1")High |
Display of Last Reading |
Solid Filled or Hollow Digits Indicate Coupled or Uncoupled Condition |
Setups |
9 Standard Setups for Probes |
Alarm Settings |
Minimum and Maximum Alarms, Range of 0.25 mm to 508 mm, 0. (0.010" to 20.00") Reading Alternates Between Solid and Hollow When Alarms Are Enabled and Violated |
Power Requirements |
2 “AA" Size Batteries |
Battery Life/Operating Time |
Approximately 60 Hours |
Instrument Shut-0ff |
Selectable ALWAYS ON or AUTO OFF After 5, 10, 15, 30 Minutes of Inactivity |
Language |
Selectable English, German, French, Spanish, Italian, Russian, Japanese and Chinese |
I/O Connectors |
Transducer |
Dual Lemo 00 (coax) |
Mini-USB |
Mini-USB to PC |
Temperature |
Operating |
-10°C to +50°C (+10°F to +120°F) |
Storage |
-20°C to +60°C (-10°F to +140°F) |
Weight |
223 g (0.597 lb) Including Batteries |
Size |
138 mm x 32 mm x 75 mm |
Shock |
IEC 68-2-27 Ea, as per Mil Std 810C Method 516.2 Procedure I with a 15g 11ms Impulse Half Sinusoidal Wave Applied 6 Times per Axis |
Sealing |
IEC529 / IP54, Dust Proof/Dripping Water Proof as per IEC 529 Specifications for IP54 Classification |
|
Data Recorder Option Features |
Capacity |
50,000 Readings |
File Structures |
Grid File |
Number of Rows |
1 to 50,000 |
Number of Columns |
1 to 223 |
File Naming |
Up to 24 Character Alphanumeric Name |
Optional Software |
UltraMATE and UltraMATE Lite |
DM5E Probe/Transducer Specifications
|
Model DA501 |
Model DA503 |
Model DA512 |
Model DA590 |
Frequency |
5 MHz |
2 MHz |
7.5 MHz |
5 MHz |
Probe Style |
Standard |
Standard |
Fingertip |
High-Temperature |
Operating Temperature Range (continuous) |
-10 to 70°C |
-10 to 70°C |
-20 to 70°C |
-20 to 204°C |
Contact Diameter |
12 mm (0.470") |
16.1 mm (0.630") |
7.5 mm (0.300") |
12.7 mm (0.500") |
IP to First Measurement Range |
1.0 to 200 mm (0.040 to 8") |
5.0 to 300 mm (0.200 to 12") |
0.6 to 60 mm (0.024 to 2.4") |
1.0 to 125 mm @20°C (0.040 to 5" @68°F)
1.3 to 25.4 mm @204°C (0.050 to 1" @400°F) |
Minimum Multi-Echo Measurement Range |
3.0 to 25 mm (0.120 to 1.000") |
N/A |
2.0 to 10 mm (0.080 to 0.400") |
N/A |
|